Quality control coding: Difference between revisions

From Atomix
No edit summary
No edit summary
 
(7 intermediate revisions by 2 users not shown)
Line 1: Line 1:
'''In progress
'''[In progress]
'''
'''


Line 5: Line 5:
'''Shear-probe quality-control flags'''
'''Shear-probe quality-control flags'''


The Q (quality control) flags associated with shear-probe measurements are not compatible with the Ocean Sites [http://www.oceansites.org/ Ocean Sites] for quality control (QC) coding.
The Q (quality control) flags associated with shear-probe measurements are bitwise flags with boolean values which are CF compliant. More information available here: https://mplnet.gsfc.nasa.gov/about-flags


Every dissipation estimate from every probe must have Q flag.
Every dissipation estimate from every probe must have Q flag.
Line 16: Line 16:
! Flag Meaning
! Flag Meaning
! Example threshold value
! Example threshold value
! Ex: True =1  / False =0  
| Ex: True =1  / False =0  
! Ex: Q value
| Ex: Q value
|-
|-
| 1
| 1
| Bit 0
| Bit 0
| if FOM > FOM_limit
| Poor figure of merit
| 2
| 1.15
| 0
| 0
| 0
| 0
Line 28: Line 28:
| 2
| 2
| Bit 1
| Bit 1
| if despike_fraction > despike_fraction_limit
| Large fraction of data with spikes
| 40%
| 5%
| 0
| 0
| 0
| 0
Line 35: Line 35:
| 4
| 4
| Bit 2
| Bit 2
| if |log(e_max)-log(e_min)|> diss_ratio_limit X \sigma_{\ln\varepsilon}
| Anomalously large disagreement between dissipation estimates from probes
| N/A
| 2.772
| 1
| 1
| 4
| 4
Line 42: Line 42:
| 8
| 8
| Bit 3
| Bit 3
| if despike_iterations > despike_iterations_limit
| Too many iterations of despiking routine
| To be confirmed
| 8
| 0
| 0
| 0
| 0
Line 49: Line 49:
| 16
| 16
| Bit 4
| Bit 4
| if variance resolved less than a threshold
| Insufficient variance resolved
| 50%
| 50%
| 1
| 1
Line 75: Line 75:
| 0
| 0
|-
|-
|
|
|
|
|
| Final Q = 20
| Final Q = 20
|}
|}
Line 85: Line 90:
The reasons for a failure can be decoded by breaking the value of Q down to its powers of 2.
The reasons for a failure can be decoded by breaking the value of Q down to its powers of 2.


-----------------------------
return to [[Flow chart for shear probes]]


[[Category: Shear probes]]
[[Category: Shear probes]]

Latest revision as of 09:18, 28 August 2023

[In progress]


Shear-probe quality-control flags

The Q (quality control) flags associated with shear-probe measurements are bitwise flags with boolean values which are CF compliant. More information available here: https://mplnet.gsfc.nasa.gov/about-flags

Every dissipation estimate from every probe must have Q flag. The numerical values of the Q flags are as follows:

Flag Mask Bit Flag Meaning Example threshold value Ex: True =1 / False =0 Ex: Q value
1 Bit 0 Poor figure of merit 1.15 0 0
2 Bit 1 Large fraction of data with spikes 5% 0 0
4 Bit 2 Anomalously large disagreement between dissipation estimates from probes 2.772 1 4
8 Bit 3 Too many iterations of despiking routine 8 0 0
16 Bit 4 Insufficient variance resolved 50% 1 16
32 Bit 5 manual flag to be defined by user N/A 0 0
64 Bit 6 manual flag to be defined by user N/A 0 0
128 Bit 7 manual flag to be defined by user N/A 0 0
Final Q = 20


The Q flags are combined by their addition. For example a Q value of 20 means that the dissipation estimated failed both dissipation ratio limit test and the resolved variance test. A value of 255 means that all tests failed. The reasons for a failure can be decoded by breaking the value of Q down to its powers of 2.


return to Flow chart for shear probes