Dataset requirements for shear probes: Difference between revisions

From Atomix
Aleboyer (talk | contribs)
No edit summary
Aleboyer (talk | contribs)
No edit summary
Line 26: Line 26:


== Recommended data format ==
== Recommended data format ==
<div class="mw-collapsible mw-collapsed" id="Recommended" data-collapsetext="Collapse" data-expandtext="data format">
<div class="mw-collapsible mw-collapsed" id="Recommended" data-collapsetext="Collapse" data-expandtext="Expand data format">


For sharing a data set, we recommend to provide four processing levels of data in a single netCDF file per instrument file. The processing levels include  
For sharing a data set, we recommend providing four processing levels of data in a single netCDF file per instrument file. The processing levels include  


* L1_converted ([[Level 1 data (shear probes)|level 1]])
* L1_converted ([[Level 1 data (shear probes)|level 1]])
Line 45: Line 45:
<div class="mw-collapsible mw-collapsed" id="raw" data-collapsetext="Collapse" data-expandtext="Expand Raw">
<div class="mw-collapsible mw-collapsed" id="raw" data-collapsetext="Collapse" data-expandtext="Expand Raw">


[[User:CynthiaBluteau|CynthiaBluteau]] ([[User talk:CynthiaBluteau|talk]]) 19:44, 28 October 2021 (CEST) COuld be collapsible tables or simply bullet lists. Instead of each sublevel into subheadings, they can be in their separate pages.
==Dimensions==
==Dimensions==
'''Example of collapsible list'''
'''Example of collapsible list'''

Revision as of 03:42, 5 November 2021

The absolute minimum data set

The absolute minimum signals and information required to estimate the rate of dissipation of kinetic energy consists of:

  • a sampled shear-probe signal,
  • a means of determining the speed of flow past the shear probe, and
  • a means to estimate the temperature of the fluid so that you can determine its kinematic viscosity.

The sampling rate of the shear-probe signal should exceed significantly (by a factor of 2 to 4) the value of 150U, where U is the speed of the flow past the shear-probe. This speed is often called the speed of profiling, even when the probe is fixed in space and the current causes fluid to flow past the shear probe. The value of 150 cpm equals the spatial resolution of the commonly used shear probe.


The desirable data set

A far more desirable, but not absolutely necessary, set of data consists of the following.

  • Two or more shear probes for measurement redundancy.
  • Vibration sensors or accelerometers that are sensitive to the directions of sensitivity of the shear probes.
  • An accurate thermometer.
  • A pressure transducer.
  • A signal that can be used to estimate the speed of flow past the shear probe.
  • Pitch and roll sensors.

Most commonly available instruments carry two shear probes that are frequently oriented so that they measure orthogonal components of the shear and, thus, provide two independent estimates of the rate of dissipation that you should agree statistically.