Level 3 data (shear probes)

From Atomix
Revision as of 12:40, 27 March 2022 by Ilker (talk | contribs)

Level 3 data (shear probes)

Level 3 data contains the raw and cleaned spectra derived from level 2 data. Level 3 parameters are defined along a new TIME dimension, which is the average time of the individual spectral segments. The length of the TIME dimension equals the number of spectral segments. Parameter for the calculation of the spectra, e.g. segment length, are provided in the meta data.

Parameter Name Standard Name Units of measurement dimension Comment
FLOWPS flow_past_sensor m s-1 TIME
PRES water_pressure dbar TIME
TIME time days since 1950-01-01 00:00:00 UTC TIME
TEMP water_temperature degree_Celsius TIME
SECTION_NUMBER dimensionless TIME
SH_SPEC shear_probe_spectrum s-2 cpm-1 [TIME, WAVENUMBER, N_SHEAR_SENSORS] from shear probes (dirty)
SH_SPEC_CLEAN shear_probe_spectrum_clean s-2 cpm-1 [TIME, WAVENUMBER, N_SHEAR_SENSORS] cleaned using Goodman algorithm
A_SPEC acceleration_sensor_spectrum m2 s-4 cpm-1 [TIME, WAVENUMBER, N_ACCEL_SENSORS]
AP_SPEC vibration_sensor_spectrum NA [TIME, WAVENUMBER, N_VIB_SENSORS]
SH_A_SPEC shear_and_acceleration_cross-spectral_matrix [TIME, WAVENUMBER, N_ACCEL_SENSORS*N_SHEAR_SENSORS]
SH_AP_SPEC shear_and_vibration_cross-spectral_matrix [TIME, WAVENUMBER, N_VIB_SENSORS*N_SHEAR_SENSORS]
KCYC Cyclic wavenumber cpm WAVENUMBER
KMAX maximum_wavenumber cpm TIME maximum wavenumber for integration of shear spectrum

Return Dataset requirements for shear probes.