{{DefineConcept
|description=Spatial averaging effect of the tip of a shear probe
|article_type=Concepts
}}
The probe-response correction depends on the dimensions of the shear probe.  
The probe-response correction depends on the dimensions of the shear probe.  
For a commonly used shear probe the spectrum is corrected for the spatial averaging of the shear probe by multiplying it by the factor correction <math>1+\left( \frac{k}{k_0} \right)^2</math> where <math>k_0</math> = 50 cpm and <math>k</math> is the wavenumber expressed in units of cpm  
For a commonly used shear probe the spectrum is corrected for the spatial averaging of the shear probe by multiplying it by the factor <math>1+\left( \frac{k}{k_0} \right)^2</math> where <math>k_0</math> = 50 cpm and <math>k</math> is the wavenumber expressed in units of cpm  
<ref name="macounlueck2004">{{Cite journal
<ref name="macounlueck2004">{{Cite journal
|authors= P. Macoun and R. Lueck
|authors= P. Macoun and R. Lueck
Line 16: Line 21:
----------------------------
----------------------------
return to [[Flow chart for shear probes]]
return to [[Flow chart for shear probes]]
[[Category:Shear probes]]

Latest revision as of 16:53, 7 May 2022


Short definition of Wavenumber response of the shear probe
Spatial averaging effect of the tip of a shear probe

This is the common definition for Wavenumber response of the shear probe, but other definitions maybe discussed within the wiki.


The probe-response correction depends on the dimensions of the shear probe. For a commonly used shear probe the spectrum is corrected for the spatial averaging of the shear probe by multiplying it by the factor [math]\displaystyle{ 1+\left( \frac{k}{k_0} \right)^2 }[/math] where [math]\displaystyle{ k_0 }[/math] = 50 cpm and [math]\displaystyle{ k }[/math] is the wavenumber expressed in units of cpm [1] The correction reaches a factor of 10 at a wavenumber of 150 cpm, and it is not recommended to use spectral data beyond this wavenumber unless your shear probe is small enough to have a cutoff wavenumber [math]\displaystyle{ k_0 }[/math] that is larger than 50 cpm.

References

  1. P. Macoun and R. Lueck. 2004. Modeling the spatial response of the airfoil shear probe using different sized probes. J. Atmos. Oceanic Technol.. doi:10.1175/1520-0426(2004)021