The probe-response correction depends on the dimensions of the shear probe.  
The probe-response correction depends on the dimensions of the shear probe.  
For a commonly used shear probe the spectrum is corrected for the spatial averaging of the shear probe by multiplying it by the factor correction <math>1+\left( \frac{k}{k_0} \right)^2</math> where <math>k_0</math> = 50 cpm and <math>k</math> is the wavenumber expressed in units of cpm  
For a commonly used shear probe the spectrum is corrected for the spatial averaging of the shear probe by multiplying it by the factor <math>1+\left( \frac{k}{k_0} \right)^2</math> where <math>k_0</math> = 50 cpm and <math>k</math> is the wavenumber expressed in units of cpm  
<ref name="macounlueck2004">{{Cite journal
<ref name="macounlueck2004">{{Cite journal
|authors= P. Macoun and R. Lueck
|authors= P. Macoun and R. Lueck

Revision as of 22:41, 19 November 2021

The probe-response correction depends on the dimensions of the shear probe. For a commonly used shear probe the spectrum is corrected for the spatial averaging of the shear probe by multiplying it by the factor [math]\displaystyle{ 1+\left( \frac{k}{k_0} \right)^2 }[/math] where [math]\displaystyle{ k_0 }[/math] = 50 cpm and [math]\displaystyle{ k }[/math] is the wavenumber expressed in units of cpm [1] The correction reaches a factor of 10 at a wavenumber of 150 cpm, and it is not recommended to use spectral data beyond this wavenumber unless your shear probe is small enough to have a cutoff wavenumber [math]\displaystyle{ k_0 }[/math] that is larger than 50 cpm.

References

  1. P. Macoun and R. Lueck. 2004. Modeling the spatial response of the airfoil shear probe using different sized probes. J. Atmos. Oceanic Technol.. doi:10.1175/1520-0426(2004)021