Line 5: Line 5:
<math> s=\frac{N_s}{2\sqrt2SU^2G\gamma} </math>
<math> s=\frac{N_s}{2\sqrt2SU^2G\gamma} </math>


where <math>s</math> is the shear signal in physical units of <math>\mathrm{s}^{-1}</math> , <math>N_s</math> are the raw numeric samples (the output of an analog-to-digital converter), <math>S</math> is the calibrated sensitivity of the shear-probe in units of <math>\mathrm{V/(m\,s^{-1})^2}</math>, <math>U</math> is the speed of profiling in <math>\mathrm{m\,s^{-1}}</math>, <math>G</math> is the gain of the electronics of the shear probe in units of <math>\mathrm{s}</math>, and <math>\gamma</math> is the gain of the analog-to-digital converter used to create the samples in units of <math>\mathrm{counts\, V^{-1}}</math>.
where <math>s</math> is the shear signal in physical units of <math>\mathrm{s}^{-1}</math> , <math>N_s</math> are the raw numeric samples (the output of an analog-to-digital converter), <math>S</math> is the calibrated sensitivity of the shear-probe in units of <math>\mathrm{V/(m\,s^{-1})^2}</math>, <math>U</math> is the speed of profiling in <math>\mathrm{m\,s^{-1}}</math>, <math>G</math> is the gain of the differentiator in the electronics of the shear probe in units of <math>\mathrm{s}</math>, and <math>\gamma</math> is the gain of the analog-to-digital converter used to create the samples in units of <math>\mathrm{counts\, V^{-1}}</math>.
The above formula assumes that the electronics of the shear probe uses a differentiator to produce a signal that is proportional to the time rate of change of the cross-profile velocity fluctuations.  


The sensitivity of the shear probe to shear is proportional to the square of the speed of profiling.  
The sensitivity of the shear probe to shear is proportional to the square of the speed of profiling.  
Thus, one should set a minimum speed for the conversion of the shear-probe data into physical units. Otherwise, the conversion may produce enormously large and quite unrealistic values.  
Thus, one should set a minimum speed for the conversion of the shear-probe data into physical units. Otherwise, the conversion may produce enormously large and quite unrealistic values.  
Realistic minimum speeds for the conversion into physical units are <math>0.05</math> to <math>0.1\, \mathrm{m\,s^{-1}}</math> because the shear-probe signal is likely to be dominated by electronic noise and the angle of attack will be large (<math> >20^{\circ}</math>) even for low levels of dissipation (\citet{Lueck, R.G., D. Huang, D. Newman, and J. Box, 1997, Turbulence measurements with a moored instrument, J. Atmos. Oceanic Techno., 14, 143-161}).
Realistic minimum speeds for the conversion into physical units are <math>0.05</math> to <math>0.1\, \mathrm{m\,s^{-1}}</math> because the shear-probe signal is likely to be dominated by electronic noise and the angle of attack will be large (<math> >20^{\circ}</math>) even for low levels of dissipation (\citet{Lueck, R.G., D. Huang, D. Newman, and J. Box, 1997, Turbulence measurements with a moored instrument, J. Atmos. Oceanic Techno., 14, 143-161}).
If your instrument does not have a differentiator, then the recorded signal is proportional to the cross-profile velocity fluctuations, say <math>u</math>.
You must convert this signal into a the time derivative of the rate of change of cross-profile velocity using the first difference operator using
<math> s = f_s\, \left[ u(n) - u(n-1) \right]  </math>
where <math>f_s</math> is the sampling rate of your data and <math>u</math> are the samples of the shear probe that have been converted in velocity fluctuations in units of <math>\mathrm{m\, s^{-1}}</math>.

Revision as of 18:55, 13 July 2021

copy&paste from V1 docx

Convert the shear-probe data samples into physical units using the standard equation,

[math]\displaystyle{ s=\frac{N_s}{2\sqrt2SU^2G\gamma} }[/math]

where [math]\displaystyle{ s }[/math] is the shear signal in physical units of [math]\displaystyle{ \mathrm{s}^{-1} }[/math] , [math]\displaystyle{ N_s }[/math] are the raw numeric samples (the output of an analog-to-digital converter), [math]\displaystyle{ S }[/math] is the calibrated sensitivity of the shear-probe in units of [math]\displaystyle{ \mathrm{V/(m\,s^{-1})^2} }[/math], [math]\displaystyle{ U }[/math] is the speed of profiling in [math]\displaystyle{ \mathrm{m\,s^{-1}} }[/math], [math]\displaystyle{ G }[/math] is the gain of the differentiator in the electronics of the shear probe in units of [math]\displaystyle{ \mathrm{s} }[/math], and [math]\displaystyle{ \gamma }[/math] is the gain of the analog-to-digital converter used to create the samples in units of [math]\displaystyle{ \mathrm{counts\, V^{-1}} }[/math]. The above formula assumes that the electronics of the shear probe uses a differentiator to produce a signal that is proportional to the time rate of change of the cross-profile velocity fluctuations.

The sensitivity of the shear probe to shear is proportional to the square of the speed of profiling. Thus, one should set a minimum speed for the conversion of the shear-probe data into physical units. Otherwise, the conversion may produce enormously large and quite unrealistic values. Realistic minimum speeds for the conversion into physical units are [math]\displaystyle{ 0.05 }[/math] to [math]\displaystyle{ 0.1\, \mathrm{m\,s^{-1}} }[/math] because the shear-probe signal is likely to be dominated by electronic noise and the angle of attack will be large ([math]\displaystyle{ \gt 20^{\circ} }[/math]) even for low levels of dissipation (\citet{Lueck, R.G., D. Huang, D. Newman, and J. Box, 1997, Turbulence measurements with a moored instrument, J. Atmos. Oceanic Techno., 14, 143-161}).

If your instrument does not have a differentiator, then the recorded signal is proportional to the cross-profile velocity fluctuations, say [math]\displaystyle{ u }[/math]. You must convert this signal into a the time derivative of the rate of change of cross-profile velocity using the first difference operator using

[math]\displaystyle{ s = f_s\, \left[ u(n) - u(n-1) \right] }[/math]

where [math]\displaystyle{ f_s }[/math] is the sampling rate of your data and [math]\displaystyle{ u }[/math] are the samples of the shear probe that have been converted in velocity fluctuations in units of [math]\displaystyle{ \mathrm{m\, s^{-1}} }[/math].