Convert the shear-probe data samples into physical units using the standard equation,
where is the shear signal in physical units of , are the raw numeric samples (the output of an analog-to-digital converter), is the calibrated sensitivity of the shear-probe in units of , is the speed of profiling in , is the gain of the electronics of the shear probe in units of , and is the gain of the analog-to-digital converter used to create the samples in units of .
The sensitivity of the shear probe to shear is proportional to the square of the speed of profiling.
Thus, one should set a minimum speed for the conversion of the shear-probe data into physical units. Otherwise, the conversion may produce enormously large and quite unrealistic values.
Realistic minimum speeds for the conversion into physical units are to because the shear-probe signal is likely to be dominated by electronic noise and the angle of attack will be large () even for low levels of dissipation (\citet{Lueck, R.G., D. Huang, D. Newman, and J. Box, 1997, Turbulence measurements with a moored instrument, J. Atmos. Oceanic Techno., 14, 143-161}).