Data processing of raw measurements: Difference between revisions
From Atomix
Created page with "The following are non-instrument specific recommendations for processing of raw measurements. As such, an initial removal of bad data based on recommendations of the instrumen..." |
No edit summary |
||
| (20 intermediate revisions by 2 users not shown) | |||
| Line 1: | Line 1: | ||
The following are | {{ADV processing | ||
|instrument_type=ADV | |||
|level=level 1 raw | |||
}} | |||
The following are specific recommendations for processing raw measurements, which are non-specific to a brand/make of instrument. As such, initial removal of bad data based on recommendations of the instrument manufacturer should be performed first (e.g. removing data with low correlation). Following that, you must: | |||
* | * [[Velocity despiking|Despike the velocity observations]] | ||
* | * [[Phase unwrapping]] if required | ||
---- | |||
Return to [[Flow chart for velocity point-measurements]] | |||
Continue to [[Preparing quality-controlled velocities]] | |||
Latest revision as of 21:31, 8 July 2026
The following are specific recommendations for processing raw measurements, which are non-specific to a brand/make of instrument. As such, initial removal of bad data based on recommendations of the instrument manufacturer should be performed first (e.g. removing data with low correlation). Following that, you must:
- Despike the velocity observations
- Phase unwrapping if required
Return to Flow chart for velocity point-measurements
Continue to Preparing quality-controlled velocities