The following are specific recommendations for processing raw measurements, which are non-specific to a brand/make of instrument. As such, initial removal of bad data based on recommendations of the instrument manufacturer should be performed first (e.g. removing data with low correlation). Following that, you must: | The following are specific recommendations for processing raw measurements, which are non-specific to a brand/make of instrument. As such, initial removal of bad data based on recommendations of the instrument manufacturer should be performed first (e.g. removing data with low correlation). Following that, you must: | ||
* Despike the data | * [[Velocity despiking|Despike the data]] | ||
* Remove segments of the time series where there is [[interference from the instrument frame]] | * Remove segments of the time series where there is [[interference from the instrument frame]] | ||
* Determine the noise levels | * Determine the noise levels |
Revision as of 18:51, 29 October 2021
The following are specific recommendations for processing raw measurements, which are non-specific to a brand/make of instrument. As such, initial removal of bad data based on recommendations of the instrument manufacturer should be performed first (e.g. removing data with low correlation). Following that, you must:
- Despike the data
- Remove segments of the time series where there is interference from the instrument frame
- Determine the noise levels